کیت
Solarbio

Serum Total Iron Binding Capacity(TIBC) Assay Kit

توضیحات و پروتکل‌های فنی

Determination of Significance: Total iron-binding capacity (TIBC) refers to the ability of serum transferrin to bind iron, and its content is closely related to the diseases such as iron deficiency anemia and acute hepatitis. Measurement Principle: Fe 2+ reacts with ferrozine to form a fuchsia compound which has an absorption peak at 562nm. In alkaline condition, serum transferrin can bind with Fe 3+ , and the remaining unbound Fe 3+ can be reduced to Fe 2+ . So the absorbance A1 is positively correlated with Fe 3+ . After acidification, the transferrin-bound Fe 3+ is released and further reduced to Fe 2+ . The absorbance A2 has a positive correlation with Fe 3+ , A2 minus A1 was proportional to TIBC. Self Provided: Spectrophotometer/microplate reader, water bath/ constant temperature foster box, centrifuge, micro glass cuvette/ 96 well plate, distilled water. Note: Product information may be optimized and upgraded. Please refer to the actual label information for accuracy. Note: Product information may be optimized and upgraded. Please refer to the actual label information for accuracy.
مشخصات و اطلاعات آزمایشگاهی
کد کاتالوگ:
C-BC2865-100T
وضعیت تامین:
موجود در انبار سیترا بیوتک
Storage:
Store at 2-8℃ 6 months
Detection Method:
Spectrophotometer/Microplate Reader
Wavelength:
562nm
تحویل:
یک ماه پس از سفارش
24000000
24000000 تومان